Paul Maierhofer 1, Georg Röhrer2, Markus Bainschab1, Alexander Bergmann1


Institute of Electronic Sensor Systems, Graz University of Technology, 8010 Graz, Austria
ams AG, 8141 Premstätten, Austria


Received: January 30, 2019
Revised: May 9, 2019
Accepted: May 11, 2019
Download Citation: ||https://doi.org/10.4209/aaqr.2019.01.0048  


Cite this article:

Maierhofer, P., Röhrer, G., Bainschab, M. and Bergmann, A. (2020). On the Inherent Variability of Particulate Matter Concentrations on Small Scales and the Consequences for Miniaturized Particle Sensors. Aerosol Air Qual. Res. 20: 271-280. https://doi.org/10.4209/aaqr.2019.01.0048


HIGHLIGHTS

  • A model for the variability of particle concentrations on small scales is proposed.
  • The variability for sample volumes of milliliters can be tens of percent.
  • Inherent variability of PM concentrations affects miniaturized sensors uncertainty.
  • Expected measurement uncertainties are shown.
 

ABSTRACT


Despite all the evident benefits of miniaturized particulate matter (PM) sensors, an inherent drawback exists in the uncertainty and validity of the measurement, which is closely related to the discrete nature of particulates suspended in air. The miniaturization of these devices not only leads to a smaller footprint for the devices themselves but also to a smaller volume of air being sampled. Even if a perfect measurement system is assumed, an uncertainty lies in assigning a supposedly representative particle concentration value to an environment due to the inherent variability of PM concentrations on small scales. This stems from the fact that particles are stochastically distributed in the air, leading to a non-uniform concentration for arbitrarily small volumes. Consequently, an uncertainty exists according to counting statistics, as the number of investigated particles in a small air sample is also low. Depending on the metric, the uncertainty may be augmented, as a small number of particles cannot accurately capture the distribution of particle sizes, especially since the size distribution extends over several orders of magnitude. This distribution related uncertainty is relevant for surface and mass related metrics in addition to the uncertainty resulting from counting statistics. We detected a minor impact from the distribution of the particle mass density, which contributes to the uncertainty for mass-related metrics, such as PM1, PM2.5 and PM10. We investigated the expected measurement uncertainty by analytical means and concluded that the distribution of particle sizes, the sample size and the ambient particle concentration significantly affect the measurement uncertainty for the range of conditions considered. To the best of our knowledge, this uncertainty has not been discussed in the current literature.


Keywords: Aerosol sampling; Measurement uncertainty; Miniaturized sensors.


INTRODUCTION


Serious adverse health effects of particulate matter (PM) on the human body (Ranft et al., 2009; Kioumourtzoglou et al., 2016) raise a public desire for highly integrated, cost-effective and easy-to-use PM sensors. As the chase for smaller and smaller PM sensors is currently at an all-time high (Paprotny et al., 2013; Li et al., 2014; Ciccarella et al., 2016; Bao et al., 2018; Hapidin et al., 2019), it is not only fruitful but also necessary to have a closer look at limitations we encounter. The miniaturization of microfluidic devices, such as miniaturized PM sensors, yields lower flow rates compared to available macroscopic devices, e.g., Papronty et al. (2013) report a flow rate of only 6 mL min–1. Therefore, the sample size of PM sensors is typically decreasing for a given integration time as they get smaller. As the sampled volume gets smaller, also the number of particles analyzed and consequently used to estimate ambient PM concentrations gets lower.

The target of this work is to evaluate the impact of miniaturization on the validity of the measurements, neglecting all possible imperfections of a sensor. The uncertainty of the estimated ambient particle concentration given the measurement of a sensor is equal to the inherent variability of PM on the scale of the sample volume. We find three separate contributions to the total uncertainty: One stems from counting statistics, the second one is caused by the broad distribution of particle sizes in the ambient environment, which plays a role for metrics such as surface concentration and mass concentration, and a third contribution to the total measurement uncertainty comes from the distribution of particle mass densities. A small number of particles does not represent this complex situation ideally. The smaller the sample gets, the higher the variability between measurements within the same ambient conditions will be. This uncertainty is also dependent on the metric used, which can depend on the sensor effect in use.

The discussed uncertainty is not covered by the current literature to the best of our knowledge.


METHODS


We state the analytical expressions of three investigated metrics in terms of the relevant particle properties as shown in Eq. (1). We use these analytical expressions to find the expectation value and the variance of the metrics on small volumes.

We can write the particulate concentrations within a sample volume as PN, PS and PMx, as in Eq. (1). PN is the particle number concentration. PS corresponds to the total particle surface area per volume. PMx is a placeholder for the most used mass metrics, such as PM1, PM2.5 or PM10. These metrics count the total mass of particles up to a diameter indicated after “PM” in units of µm. For example, PM1 corresponds to the total mass of particles less than 1 µm in diameter per cubic meter.

where N is the total number of particles within the sample, and Sn, Vn and ρn are the surface, volume and the mass density of the nth particle, respectively. VSample denotes the sampled volume of air. Mathematically speaking, N is a random positive integer number, and Sn, Vn and ρn are each independent and identically distributed random variables, which means that, e.g., all Vn have the same probability distribution, with finite expectation value and variance.

For the particle number concentration, only the number of particles per volume is relevant, whereas for the surface concentration, the individual surfaces of the single particles are summed up and for mass concentrations, their masses. This concentration will, especially for small sample volumes, vary from the mean ambient particle concentration since the suspended particles are discrete objects. In an extreme case, we could think of a situation where only one particle is suspended in a known volume of air. We can assume a sensor, which would need a large number (m) of measurements to analyze the whole air in the total volume. This sensor would report in m – 1 measurements a concentration of 0 and in one measurement, a concentration of m times the mean concentration in said volume. Clearly, this sensor would not be able to capture the situation with only one measurement—not because of imperfections of the sensor itself but because of the high variability in the particle concentration. Another effect stems from the size distribution and the mass density distribution of particles, which complicates the situation for surface concentration and mass concentration metrics.

A sensor will use the measurement value of the sampled volume to estimate the particle concentration on a larger scale. Therefore, we are interested in the expectation value, < >, and the variance, Var(), for each of the investigated metrics, which we calculate using Wald’s equation as well as Blackwell-Girshick’s equation (Klenke, 2013) and the assumption that the number of particles (N), the surface (S), the volume (V) and the density (ρ) are uncorrelated. 

Eq. (3) shows the total variance of each metric’s value. For PN, the situation is straightforward since only the counting uncertainty plays a role. The surface concentration has an additional term including the variance of a single particle’s surface. PMx has a term proportional to the variance of a single particle’s volume, Var(V), as well as a term with the variance of the mass density, Var(ρ). Var(V) and Var(S) are functions of the particle-size distribution. This distribution typically ranges in terms of diameter over decades, from nanometers up to micrometers, and is not ideally captured by a small number of particles; therefore, it is obvious that the size distribution has an influence on the total uncertainty. In order to evaluate the variances, a discussion of the distributions of the number of particles (N), the surface and the volume of particles (S and V) and the particle density (ρ) is necessary. Since both surface and volume distribution of the particles are functions of the size distribution, we can discuss the more frequently used particle size distribution.


Particle Size Distribution

The size distribution of particles is frequently assumed to be lognormal (Hinds, 1999) in diameter, D, but can, in principle, have any shape. In order for Eqs. (2)–(3) to be valid, the variances of S and V must be finite, which is the case for any realistic distribution since diameter, surface and volume of particles are always finite. Since the size distribution of particles is not constant in the ambient air but depends on season, weather, nearby emissions and a variety of other environmental parameters (e.g., Väkevä et al., 2000), it is not possible to assume one size distribution that is representative for all cases.

Given the size distribution, p(D), as a function of the particles’ diameter, D, we can write the expectation value and the variance of a variable being dependent on D, such as the surface (S) and the volume (V), as in Eqs. (4)–(5), where V is used as an example.

Eqs. (4)–(5) show the expectation value, <V>, and the variance, Var(V), of a single particle’s volume, which is only dependent on the distribution p(D). Since the variance of S and  is a function of this distribution, we can see that the distribution indeed affects the variability of the mass and the surface concentration of particulate matter directly. In order to evaluate Eqs. (4)–(5), a particle shape has to be assumed in order to find surface and volume as a function of diameter. The model is valid for any particle shape. Without any loss of generality, we choose spherical particle shapes for further evaluations for the sake of simplicity.


Distribution of Particle Mass Density

The distribution of the mass density, ρ, is—similar to the size distribution—dependent on various factors, such as the type of particle sources near the measurement. Measurement principles that do not directly determine mass—such as light scattering—need an assumption for the mean mass density to determine PMx. An incorrect assumption of <ρ> directly affects the measured mass concentration, causing a systematic error. For the estimation of the total variability, we assume the mass density to be Gaussian distributed with a mean of <ρ> and a variance of Var(ρ). If these cannot be experimentally evaluated, the parameters have to be estimated by using prior knowledge, such as knowledge of the nearby aerosol sources. Aerosol particles, in principle, may consist of any material, which implies a broad range of mass densities theoretically possible, ranging from a few hundred kg m3 up to over 19,000 kg m–3 for metallic particles. However, for realistic ambient cases, we can typically narrow down dominant particle densities. For example, Hand and Kreidenweis (2002) report a mass density of 1850 ± 140 kg m–3 for PM2.5 measured in the southwest of Texas in the U.S.

 
Particle Number

For consistency, we require that <PN> = cPN, <PS> = cPS and <PMx> = cPMx, which means that the expectation value of each metric is equal to the respective ambient particle concentration on a large scale. We can use Eq. (2) to express the expectation value of the number of particles counted, <N>, as stated in Eq. (6). While all three equations are valid, it makes sense to use only the equation with the corresponding ambient concentration of each metric for further evaluations. 

We further assume uncorrelated particle positions, leading to uniformly distributed particles in space. Therefore, the total number of particles in a given volume will be Poisson distributed (Badger, 1946; Hinds, 1999). Although the assumption that particle positions are uncorrelated is under debate (Damit et al., 2014; Larsen, 2006), we draw the conclusion that the Poisson assumption is a valid starting point and neglect any possible correlations of particle positions, such as a clustering of particles of the same size or a complete avoidance of such clusters. In other words, we assume completely uniformly distributed particle positions. According to Larsen (2006), assuming that particles are randomly distributed in a volume, the Poisson distribution describes the number of particles within any subvolume.

Since we assume the number of particles within a sample is Poisson distributed, Eq. (7) expresses the variance of N, as the variance of the Poisson distribution is equal to its expectation value. 


Coefficient of Variation

In order to compare the uncertainties of the metrics, we use the coefficient of variation, which is the standard deviation divided by the expectation value and therefore shows the relative variability. The standard deviation of PN, PS and PMx is given by the square root of the respective variance, as given in Eq. (3). This is the inherent variability of particle concentrations on an arbitrary scale and is therefore independent of the type of PM sensor, e.g., whether the sensor will analyze single particles and add them up, or take a cumulative measurement. The type of sensor used will, of course, influence the metric of the measurement and vice versa. The coefficient of variation of the measurement can be written as stated in Eq. (8) by using Eqs. (2), (3) and (7)

In Eq. (8), we can clearly see that the coefficient of variation of each metric considered drops as the sample size increases with a typical 1/√N behaviour. As stated in Eq. (6), <N> is directly proportional to the sample volume, Vsample, and the mean ambient particle concentration in either metric. Therefore, the variability decreases for increasing sample volumes and higher concentrations, which both lead to better statistics. Another contribution to the variability of the measurement stems from the distribution of the particle surface for PS and the volume for PMx, which are both directly related to the particle size distribution. In addition, the distribution of the mass density affects the variability of mass related metrics (PMx).

The formalism can easily be adapted to metrics such as PM10, PM2.5 or PM1. The upper bound of the metric influences the expectation value and variance of the volume of single particles <V> and Var(V). For example, the variance of particle sizes of PM1 is typically lower than for PM10 because the particles of interest expand over one more decade in diameter for PM10, as 1 µm is the limit for PM1 and 10 µm for PM10. Since the volume is proportional to the diameter cubed, the variance in volume is significantly affected by the particle sizes that are taken into account. Also, the majority of particles is typically below 1 µm in diameter, which means that the Poisson counting uncertainty is not greatly affected by lowering the upper bound of the metric to 1 µm. For example, Klejnowski et al. (2013) report that during a measurement campaign in Zabrze in Poland, 99% of all particles by number concentration are below 1 µm in diameter. Given a distribution, p(D), the variance for the metric of interest can be calculated by a normalization of p(D) within the size window of interest. The probability outside this window is set to 0. Then again, Eqs. (4) and (5) can be used to calculate expectation value and variance of a single particle’s volume, respectively. Note that the ambient concentration, which is used to calculate the expectation value of the number of particles, <N>, must therefore match the metric of interest.


RESULTS AND DISCUSSION


We begin our discussion with the uncertainty of particle number concentration measurements. This uncertainty is straightforward to calculate and to estimate for a measurement, as for a PN measurement, the number of particles counted has to be known obviously. Eq. (8) can be evaluated, which gives the uncertainty. The uncertainty of the measurement can be lowered by prolonging the measurement until a relative uncertainty is reached that is acceptable for the user.

For the particle surface concentration, the situation is already more complex, as can be seen in Eq. (8). There, the expectation value and the variance of a single particle’s surface, which is a function of the particle size distribution, play a role. Depending on the measurement technique, the size distribution can be known or unknown. If there is no information on the particle size distribution, the exact uncertainty cannot be determined. Since the situation for mass concentrations, such as PMx, is similar, and these metrics are the most common metrics for low-cost sensors, we focus the discussion on mass metrics.

As shown in Eq. (8) and Fig. 1, the coefficient of variation of the measured PM2.5 value is strongly dependent on the size distribution of particles. For this evaluation, we assume the particle diameter, D, to be lognormally distributed and set ρ to 1850 ± 140 kg m–3 according to Hand and Kreidenweis (2002). For narrow particle size distributions, the variability is dominated by the Poisson contribution and the contribution from the mass density distribution. If we assume monodisperse particles, the geometric standard deviation (GSD) is 1, and Var(V) is 0. In this case, we are left with the Poisson uncertainty and the small contribution due to the mass density distribution. At broader particle size distributions with higher GSD, the size distribution related contribution dominates. Note that the standard deviations stemming from different sources do not add up, but the variances do, as shown in Eq. (3). The total coefficient of variation for the investigated case for PM2.5 ranges from roughly 4% to over 30% depending solely on the parameters of the size distribution. This major influence of the underlying particle size distribution makes the estimation of confidence intervals for a real measurement tricky. In a real case, users are forced to use the measured PMx value as an estimate for the ambient concentration. This is reasonable since the expectation value of the measurement equals the ambient concentration when there are no systematic errors in the measurement. If the measurement technique counts and sizes particles such as, for example, a light scattering device does, then Eq. (8) can be used to estimate the uncertainty. In this case, a mean mass density for the particles has to be assumed as well as a variance of the mass density. However, if the measurement technique does not provide information on the particle number and the size distribution, there is no estimate for the uncertainty of the measured PMx value as the coefficient of variation as stated in Eq. (8) cannot be evaluated. This is, for example, the case for gravimetric measurements. Also, the evaluation of Eq. (8) for a real measurement case will not lead to the exact uncertainty, as the information on the mean ambient concentration and the size distribution is not exact for finite sample sizes. In this work, we have all the information of the ambient situation and use it to find the variability of the particle concentrations. In a real case, there is little to no information on the mean ambient particle concentration, the size distribution and the mass density distribution of particles.

Fig. 1. Coefficient of variation as a function of the geometric standard deviation of the particle sizes for PM2.5.
Fig. 1. Coefficient of variation as a function of the geometric standard deviation of the particle sizes for PM2.5. GSD corresponds to the width of the distribution. The calculation is done at an exemplary geometric mean of 300 nm. For the calculation, only particles with diameters lower than 2.5 µm are taken into account in order to show the uncertainty for PM2.5. The calculation is shown for a constant ambient PM2.5 concentration of 25 µg m–3, a particle mass density <ρ> of 1850 kg m–3 with a variance Var(ρ) of (140 kg m–3)2 and a sample volume of 1 mL. Although the ambient PM2.5 concentration is held constant, and therefore, the total mass that is sampled is also constant, the total uncertainty ranges over tens of percent. For realistic cases, the uncertainty stemming from the particle size distribution dominates. The Poisson contribution also rises as GSD rises since the number of particles at a constant ambient concentration gets lower as the variance of the size distribution rises.

Mathematically speaking, the influence of the particle size distribution and the influence of the mass density distribution are equal. Nevertheless, we can state that for cases that are to be expected in the ambient air, the size distribution of particles dominates. This is reasonable since the particle size distribution ranges over more than three decades from nanometers to micrometers, while the mass density distribution can safely be assumed to extend over less than one decade. Therefore, the term Var(ρ)/<ρ>2 is, in general, small compared to Var(V)/<V>2.

Fig. 2 shows the uncertainty of PM1, PM2.5 and PM10 for lognormally distributed particles with an exemplary geometric mean (GM) of 0.3 µm as a function of the geometric standard deviation (GSD) of the size distribution. The ambient mass concentration is set to 25 µg m–3 in each of the metrics. The relative uncertainty in general rises as GSD, and therefore the width of the particle size distribution, rises. At GSD values approximately lower than 1.5, the relative uncertainties of the metrics are equal, as the metrics do not differ from each other as long as all particles are below 1 µm in diameter. For higher GSD values, the size distribution relevant to each of the metrics begins to differ, leading to a higher uncertainty of PM10 compared to PM2.5 and PM1. We can state that for the investigated parameters, the estimation of a mass concentration based on a small sample yields results that are more robust the lower the upper cutoff is. Since only a small number of particles typically has a diameter of greater than 2.5 µm but their total volume is still of relevance, PM10 is especially prone to a high uncertainty. 

Fig. 2. Coefficient of variation as a function of the geometric standard deviation of the particle sizes for PM1, PM2.5 and PM10. GSD corresponds to the width of the distribution. Fig. 2. Coefficient of variation as a function of the geometric standard deviation of the particle sizes for PM1, PM2.5 and PM10. GSD corresponds to the width of the distribution. The calculation is done at an exemplary geometric mean of 0.3 µm. The calculation is shown for an ambient particle mass concentration of 25 µg m–3 in both metrics, a particle mass density <ρ> of 1850 kg m–3 with a variance Var(ρ) of (140 kg m–3)2 and a sample volume of 1 mL. Although the ambient particle mass concentration is held constant in each metric, and therefore, the total mass that is sampled is also constant, the total uncertainty ranges over tens of percent and is a function of the upper cutoff of the relevant particle sizes. We can clearly see that at a GSD lower than ~1.5, the uncertainties are almost equal, as there are almost no particles that have a diameter above 1 µm, which makes the metrics’ output equal. At higher GSDs, the metrics begin to differ, and so do the uncertainties. The lower the upper cutoff of the metric is, the lower the Var(V) of the relevant size window is. We can see that for equal distributions and similar mass concentrations, the relative uncertainties of PM1, PM2.5 and PM10 differ, although all of them take the mass into account.

Fig. 3 shows the measurement uncertainty of PN, PS and PM2.5 as a function of the sample volume. We assumed lognormally distributed particles with an exemplary GM of 0.3 µm and a GSD of 1.7. The particle density, ρ, is set to 1850 ± 140 kg m–3 according to Hand and Kreidenweis (2002). We only took particles with diameters below 2.5 µm into account for this evaluation. In general, the relative uncertainty drops as the sample volume increases. The uncertainty of PM2.5 is the highest, followed by the uncertainty of the total particle surface (PS). The particle number concentration, PN, has the lowest uncertainty, which is immediately evident in Eq. (8), as for the uncertainty of PMx and PS, more terms are added. For lognormally distributed spherical particles, the relative uncertainty of the particle surface concentration is necessarily lower than the uncertainty of the volume concentration. The reason is that the surface is proportional to the square of the diameter of the particle, whereas the volume is proportional to the diameter cubed. This makes the mass metrics more prone to a relatively high uncertainty due to a small number of large particles. For the investigated parameters, a sample volume of 1 mL yields a relative uncertainty of 6% for PN, 11% for PS and 19% for PM2.5. Especially for PM2.5, this might be a high uncertainty, bearing in mind that it is only the inherent uncertainty of the measurement when the sensor’s imperfections are yet to be considered. For a sample volume of 1 L, the uncertainty below 1% for each metric is most likely overshadowed by the sensor’s imperfections in a realistic case, which is why this uncertainty has never been observed using high-end laboratory equipment providing high flow rates and therefore has not been discussed in literature yet.

Fig. 3. The plot shows the measurement uncertainty as a function of the sample volume. The particle size distribution is assumed to be lognormal with an exemplary GM of 0.3 µm and a GSD of 1.7. Fig. 3. The plot shows the measurement uncertainty as a function of the sample volume. The particle size distribution is assumed to be lognormal with an exemplary GM of 0.3 µm and a GSD of 1.7. The ambient PM2.5 concentration is set to 25 µg m–3 and the particle density, ρ, to 1850 ± 140 kg m–3 according to Hand and Kreidenweis (2002). σrel drops as the sample volume increases for each of the metrics. Particles with diameters higher than 2.5 µm are neglected for each metric. The relative uncertainty of PN is the lowest, followed by PS, and PM2.5 has the highest uncertainty.

In Figs. 4 and 5, we show the relative uncertainty for PM2.5 as a function of the total mass that is sampled. As the uncertainty depends on the total number of particles that are sampled, which is proportional to the sample volume times the ambient mass concentration, the total mass sampled by the sensor is a reasonable parameter. In Fig. 4, the uncertainty of PM2.5 is shown for lognormally distributed particles with a GM of 0.3 µm and various GSD values. The particle density, ρ, is set to 1850 ± 140 kg m–3 according to Hand and Kreidenweis (2002). In Fig. 5, the same is shown for a fixed GSD value of 1.5 and various values of GM. The relative uncertainty declines when higher total masses are sampled, which can be due to either a high sample volume, a high ambient particle mass concentration, or both. In general, the uncertainty rises for higher GSD values, as the size distribution becomes broader and therefore, the variance of the particles volume becomes higher. The uncertainty is also dependent on the geometric mean of the size distribution. The lower the GM, the more particles are necessary to have a fixed mass concentration.

Fig. 4. The plot shows the relative uncertainty of PM2.5 as a function of the total mass that is sampled. A lognormal particle size distribution with an exemplary GM of 0.3 µm is assumed. Fig. 4. The plot shows the relative uncertainty of PM2.5 as a function of the total mass that is sampled. A lognormal particle size distribution with an exemplary GM of 0.3 µm is assumed. The results for various GSD values are shown. Assuming a PM2.5 concentration of 25 µg m3, a volume of 4 mL is typically necessary in order to sample a total mass of 100 pg. Depending on the GSD of the particle size distribution, the relative uncertainty of PM2.5 can differ significantly.

Fig. 5. The plot shows the relative uncertainty of PM2.5 as a function of the total mass that is sampled. A lognormal particle size distribution with a GSD of 1.5 is assumed. The results for various GM values are shown. Depending on the GM, the relative uncertainty of PM2.5 can differ significantly. This is reasonable since at lower GM values, a greater number of particles is necessary to have the same mass concentration as for higher GM values. The greater number of particles lowers the relative uncertainty.

Fig. 6 shows the relative uncertainty of PM1, PM2.5 and PM10 as a function of the total mass that is sampled at an exemplary lognormal particle size distribution with a GM of 0.3 µm and a GSD of 1.7. The particle density, ρ, is set to 1850 ± 140 kg m–3 according to Hand and Kreidenweis (2002). PM1 has the lowest cutoff and therefore the lowest relative uncertainty at equal masses that are sampled, whereas PM10 has the highest of the three. 

Fig. 6. The plot shows the relative uncertainty of PM1, PM2.5 and PM10 as a function of the total mass that is sampled. For each of the metrics, the same lognormal particle size distribution with a GM of 0.3 µm and a GSD of 1.7 is assumed. Fig. 6. The plot shows the relative uncertainty of PM1, PM2.5 and PM10 as a function of the total mass that is sampled. For each of the metrics, the same lognormal particle size distribution with a GM of 0.3 µm and a GSD of 1.7 is assumed. The relative uncertainties of PM1, PM2.5 and PM10 differ due to the upper cutoff in particle size that is considered. With a total sampled mass of 100 pg, the uncertainty is 5% for PM1, 10% for PM2.5 and 11% for PM10. For PM1, 30 pg are sufficient for the assumed parameters to have a relative uncertainty below 10%.

As a rule of thumb, we propose that for PM2.5 and PM10, a minimum total mass of 100 pg should be sampled, which yields a measurement uncertainty for PM2.5 of ~10%, as can be seen in Figs. 4 and 6. This assumes lognormally distributed particles with a geometric mean of 0.3 µm, a GSD of 1.7 and a particle density (ρ) of 1850 ± 140 kg m–3. We consider this a realistic worst-case scenario. At a PM2.5 mass concentration of 25 µg m3, this would mean that approximately 4 mL of air have to be sampled; at 10 µg m3, 10 mL would be necessary to stay below the 10% measurement uncertainty. For PM1, we propose a minimum of 30 pg be sampled in order to have a relative uncertainty below 10%. We would like to emphasize that this is not valid for arbitrary particle size distributions and mass density distributions. In order to evaluate the exact uncertainty, Eq. (8) has to be utilized.

For particle number measurements, the rule of thumb we propose is to count at least 100 particles, which yields a relative uncertainty of 10%. Keeping in mind that imperfections of the measurement system itself come on top of this inherent uncertainty, a 10% inherent uncertainty may well be low compared to the uncertainties coming from technological boundaries.


CONCLUSIONS


We have established a model for estimating the local variability of particulate matter concentrations on arbitrarily small scales and evaluated it using various parameters. The model addresses the uncertainty in the particle number concentration and the surface concentration as well as the mass concentration for metrics such as PM1, PM2.5 and PM10. The variability is due to the small number of particles within a small sample, which is used to represent a rather complex distribution of particle sizes and mass densities. Although we assume that variability on a small scale has no influence on human health, it does have a major impact on the validity of miniaturized PM sensor measurements. We find that even a sensor without any imperfections is barely able to assign a reliable particle concentration to the ambient environment when the sample volume is on the order of milliliters. Depending on the metric used, the uncertainty depends on the distributions of the influencing parameters. The overall uncertainty for the particle number concentration depends on the total particle concentration and the sample volume; additionally, the particle surface concentration is dependent on the size distribution, and for mass-related metrics such as PM1, PM2.5 and PM10, also the mass density distribution has to be taken into account. These values, aside from the sample volume, may remain unknown with low-cost sensors and low sample volumes, effectively preventing the estimation of the uncertainty in the measurement. The total uncertainty is on the order of 10% for realistic cases and a sample volume of a few milliliters.

We have shown that for the same ambient conditions, the relative uncertainty in the particle number is lower than that in the surface concentration, which is lower than the relative uncertainty in the mass concentration. A small sample volume is therefore better suited to estimating number or surface concentrations than ambient mass concentrations. When mass metrics are used, the uncertainty is typically lower for estimating PM1 than PM2.5. The highest relative uncertainty is generally associated with PM10, as the range of measured particle sizes expands to include the largest diameters.

Miniaturized sensors with small footprints can overcome these limitations only by allowing relatively long measurement times or providing high flow rates. These measures, however, are incompatible with miniaturization since both features increase energy consumption and a high flow rate usually requires a large footprint. On top of these inherent limitations, sensor imperfections must be taken into account. As a rule of thumb for PM2.5 and PM10, we propose a minimum total sample mass of 100 pg—typically 4 mL of sampled air at an ambient mass concentration of 25 µg m–3—in order to cap the relative uncertainty at approximately 10% for cases we consider to be realistic for ambient measurements; to obtain a ~10% uncertainty for PM1, a minimum total sample mass of 30 pg is required. The miniaturization of particle sensors may hit a boundary once sample volumes reach ~1 mL due to the inherent statistical variability of particle concentrations on such small volumes. In the opinion of the authors, a sample volume of less than 1 mL at realistic ambient particle concentrations may be insufficient for providing any reliable results. This handicap is especially severe when measuring mass-related metrics, such as PM1, PM2.5 and PM10.


ACKNOWLEDGMENTS


This work was funded by FFG (Austrian Research Promotion Agency) Grant #86197 and ams AG. Supported by TU Graz Open Access Publishing Fund.


REFERENCES


  1. Badger, E.H.M. (1946). Particle counts in the ultramicroscope. Nature 157: 480. [Publisher Site]

  2. Bao, Y., Cai, S., Yu, H., Xu, T., Xu, P. and Li, X. (2018). A resonant cantilever based particle sensor with particle-size selection functiqn. J. Micromech. Microeng. 28: 085019. [Publisher Site]

  3. Ciccarella, P., Carminati, M., Sampietro, M. and Ferrari, G. (2016). Multichannel 65zF RMS resolution CMOS monolithic capacitive sensor for counting single micrometer-sized airborne particles on chip. IEEE J. Solid-State Circuits 51: 2545—2553. [Publisher Site]

  4. Damit, B., Wu, C. and Cheng, M. (2014). On the validity of the poisson assumption in sampling nanometer-sized aerosols. Aerosol Sci. Technol. 48: 562–570. [Publisher Site]

  5. Hand, J.L. and Kreidenweis, S.M. (2002). A new method for retrieving particle refractive index and effective density from aerosol size distribution data. Aerosol Sci. Technol. 36: 1012–1026. [Publisher Site]

  6. Hapidin, D.A., Saputra, C., Maulana, D.S., Munir, M.M. and Khairurrijal, K. (2019). Aerosol chamber characterization for commercial particulate matter (PM) sensor evaluation. Aerosol Air Qual. Res. 19: 181–194. [Publisher Site]

  7. Hinds, W.C. (1999). Aerosol Technology: Properties, Behavior, and Measurement of Airborne Particles. Wiley, New York, USA. [Publisher Site]

  8. Kioumourtzoglou, M., Schwartz, J.D., Weisskopf, M.G., Melly, S.J., Wang, Y., Dominici, F. and Zanobetti, A. (2016). Long-term PM2.5 exposure and neurological hospital admissions in the northeastern United States. Environ. Health Perspect. 124: 23–29. [Publisher Site]

  9. Klenke, A. (2013). Wahrscheinlichkeitstheorie. 3. Auflage. Springer Spektrum, Berlin-Heidelberg, Germany, pp. 103–109. [Publisher Site]

  10. Larsen, M.L. (2007). Spatial distributions of aerosol particles: Investigation of the Poisson assumption. J. Aerosol Sci. 38: 807–822. [Publisher Site]

  11. Li, X., Iervolino, E., Santagata, F., Wei, J., Yuan, C.A., Sarro, P.M. and Zhang, G.Q. (2014). Miniaturized particulate matter sensor for portable air quality monitoring devices. IEEE Sensors 2014 - Valencia Conference Centre, Valencia, Spain, pp. 2151–2154. [Publisher Site]

  12. Linden, W., Dose, V. and Toussaint, U. (2014). Bayesian Probability Theory: Applications in the Physical Sciences. Cambridge University Press, Cambridge, UK. [Publisher Site]

  13. Paprotny, I., Doering, F., Solomon, P.A., White, R.M. and Gundel, L.A. (2013). Microfabricated air-microfluidic sensor for personal monitoring of airborne particulate matter: Design, fabrication, and experimental results. Sens. Actuators, A 201: 506–516. [Publisher Site]

  14. Ranft, U., Schikowski, T., Sugiri, D., Krutmann, J. and Krämer, U. (2009). Long-term exposure to traffic-related particulate matter impairs cognitive function in the elderly. Environ. Res. 109: 1004–1011. [Publisher Site]

  15. Väkevä, M., Hämeri, K., Puhakka, T., Nilsson, E.D., Hohti, H. and Mäkelä, J.M. (2000). Effects of meteorological processes on aerosol particle size distribution in an urban background area. J. Geophys. Res. 105: 9807–9821. [Publisher Site]