Longfei Chen1, Xin Zhang1, Cuiqi Zhang1, Mohsin Raza1, Xinghua Li 2

  • 1 School of Energy and Power Engineering, Beihang University, Beijing 100191, China
  • 2 School of Space and Environment, Beihang University, Beijing 100191, China

Received: June 11, 2017
Revised: August 25, 2017
Accepted: September 13, 2017
Download Citation: ||https://doi.org/10.4209/aaqr.2017.06.0201  


Cite this article:
Chen, L., Zhang, X., Zhang, C., Raza, M. and Li, X. (2017). Experimental Investigation of a Condensation Particle Counter Challenged by Particles with Varying Wettability to Working Liquid. Aerosol Air Qual. Res. 17: 2743-2750. https://doi.org/10.4209/aaqr.2017.06.0201


HIGHLIGHTS

  • Factors influencing CPC counting efficiency were investigated for three particles.
  • The T of the saturator and condenser influences the counting efficiency of CPCs.
  • The counting efficiency of CPC is susceptible to particle wettability at small ΔT.

 

ABSTRACT


The counting efficiency of an in-house laminar flow Condensation Particle Counter (CPC) was investigated experimentally. The CPC was challenged by different sources of particles (ambient particles, and combustion particles generated from CH4 and C2H4), and the counting efficiency of the in-house ultrafine CPC (IUCPC) was obtained by comparing particle concentration data measured by the IUCPC with that measured by a reference instrument, TSI CPC 3775. The counting efficiency is closely related to the contact angle (an indication of the wettability of particles by the working fluid) and the combination of the saturator and the condenser temperatures. The particles from different sources appeared to have different contact angles, in other words, different wettability by the working liquid.


Keywords: Condensation particle counter; Counting efficiency; Wettability