Christof Asbach 1, André Schmitz2, Frank Schmidt3, Christian Monz3, Ana Maria Todea1

  • 1 Institut für Energie- und Umwelttechnik e. V. (IUTA), Air Quality & Filtration, 47229 Duisburg, Germany
  • 2 Universität Duisburg-Essen, Nanopartikel- Prozesstechnik (NPPT), 47058 Duisburg, Germany
  • 3 Institut für Gefahrstoffforschung (IGF), 44789 Bochum, Germany

Received: November 11, 2016
Revised: November 11, 2016
Accepted: January 21, 2017
Download Citation: ||https://doi.org/10.4209/aaqr.2016.10.0460  

  • Download: PDF


Cite this article:
Asbach, C., Schmitz, A., Schmidt, F., Monz, C. and Todea, A.M. (2017). Intercomparison of a Personal CPC and Different Conventional CPCs. Aerosol Air Qual. Res. 17: 1132-1141. https://doi.org/10.4209/aaqr.2016.10.0460


HIGHLIGHTS

  • Personal Enmont CPC accurately measures hydrophilic and moderately hydrophobic particles.
  • Both tested water CPCs were unable to adequately detect pure DEHS aerosols.
  • More than 20 year old TSI CPCs still measure precisely, but not accurately.

 

ABSTRACT


We here report on a study that examines the performance of a water based personal condensation particle counter (CPC) PUFP C100 (Enmont, LLC) when sampling a variety of different test aerosols. Three additional CPCs were included in the study, namely, the general purpose water based CPC model 3787 (TSI) as well as the more than 20 year old CPC model 3022 (TSI) and ultrafine CPC model 3025A (TSI), both based on butanol. The responses of these CPCs were compared to a reference ultrafine CPC model 3776 (TSI). Test aerosols included hydrophilic NaCl, moderately hydrophobic carbon and highly hydrophobic DEHS aerosols of various sizes and concentrations. The CPC PUFP C100 measured according to its specifications with high precision and accuracy as long as the particles were not highly hydrophobic, but failed to adequately measure the concentrations of highly hydrophobic DEHS aerosols. Its detection efficiency for very small particles, e.g., 5.4 nm, was decreased, which was expected based on its particle size detection limit. The TSI water CPC model 3787 showed a similar behavior for all aerosols, except that this CPC was able to accurately measure the concentration of 5.4 nm NaCl particles. The concentrations measured with the two old and unserviced butanol CPCs were highly correlated with the concentrations delivered by the reference CPC. Nevertheless, the concentrations measured with both old CPCs were clearly below the reference concentrations.


Keywords: CPC; Personal monitor; Hydrophobic; Hydrophilic


Share this article with your colleagues 

 

Subscribe to our Newsletter 

Aerosol and Air Quality Research has published over 2,000 peer-reviewed articles. Enter your email address to receive latest updates and research articles to your inbox every second week.

Latest coronavirus research from Aerosol and Air Quality Research

2020 Impact Factor: 3.063

5-Year Impact Factor: 2.857

Aerosol and Air Quality Research (AAQR) is an independently-run non-profit journal, promotes submissions of high-quality research, and strives to be one of the leading aerosol and air quality open-access journals in the world.